PERFORMANCE AND SPECIFICATIONS
AMB moisture balances provide a rapid and accurate method for moisture content
and dry weight
analysis of a wide range of products and materials. In research work, as well as in
manufacturing
and quality control applications it is essential to be able to determine moisture
content simply
and reliably. AMB infra-red moisture balances do precisely that, providing moisture
content data
as a percentage of initial mass or of the residual dry mass. The balances can also
be
set up to
display percentage solid content. A suitable sample of the product is placed inside
the drying
chamber, test parameters are entered via the keypad and the analysis is started
with
a simple
key press.
Two quartz halogen lamps provide constant, even drying of the sample at a
temperature
pre-selected by the operator. The advanced software closely monitors the
temperature in the
drying chamber and maintains it to within 1°C of the chosen setting.
As the analysis progresses the balance displays the current status, in accordance
with the
chosen parameters, keeping the operator updated on each stage of the process.
The
analysis will
automatically terminate either when drying is complete and the dry weight is stable,
or after a
fixed time specified by the operator. The final analysis result is shown on the digital
display.
At the end of the analysis all relevant data concerning initial mass, residual mass,
test
parameters and results can be printed out via an RS-232 interface. When the
moisture balance is
connected to a PC the data acquisition software allows the operator to produce a
graph plotting
results against time, providing valuable information on speed of drying and
confirmation of dry
weight stability. Results of individual tests can be stored in the computer memory
for
subsequent
comparison.
SPECIFICATIONS
Model Capacity Readability (wgt) Readability (%)
AMB 50 50g 0.001g 0.01%
AMB 110 110g 0.001g 0.01%
AMB 310 310g 0.001g 0.01%
· Temperature range: 50°C - 160°C in increments of 1°C (Optional high temperature
version)
· Verification interval settings: 2 - 59 secs in 1 second increments
· Maximum analysis time: 10 hours
· Weighing pan size: 100mm diameter
· Overall dimensions: 175mm wide × 300mm deep × 190mm high
· Gross weight: 5kg
· Power requirements: 115v (230v option)
· Heating elements: Dual quartz halogen tubes